Only use products for the maintenance of medico-surgical equipment
compatible with stainless steel (no chlorinated content).
Forbid antiseptics that are intended for use on skin and mucous.
Forbid products containing aldehyde, alcohol or other products likely to bind
proteins.
D
For each product used: refer to the manufacturer's instructions.
Comply with the concentrations, exposure durations, and life span
of products. Do not mix the products and follow the instructions for
their disposal.
W
Spraying nozzles Ref. 10476 (the green ones) cannot be sterilised:
change them after each surgery.
VI.1 – Preparation of MD in the place of use
To be done immediately after surgery (maximum time of 2 hrs):
• Disconnect the MD from the motor,
• Remove the rotary tool (see § V.6),
• Remove the largest organic contaminants with disinfectant wipes.
VI.2 – Preparation for sterilisation
Disassembly of MD (Fig 1):
• Remove the rotary instrument,
• Disconnect the MD from the motor,
• Remove the rotary instrument and the spraying nozzle.
• Pivot the F lever up to the stop (indifferently towards the right or towards
the left),
• Pull the A head,
• Remove the collar cartridge D.
VI.2.1 – Manual preparation for sterilisation
Cleaning of irrigation pipe:
Carefully clean the irrigation pipe and remove any impurities and deposits
from it using the caustic cleaner spray.
E
Risk of injury: use protective glasses.
By brushing:
• Brush each part under running water with a soft brush.
d
b
• Wipe a disinfectant wipe over each of the MD's sub-units,
• Or Spray disinfectant on the MD and wipe with a clean cloth.
20
Lighting lens (LED or Fibre optic):
• Absolutely avoid all scratches to the LED or the fibre optic,