After cleaning and before each sterilisation (once a day minimum):
• Remove the rotary instrument,
• Insert the tip to the back of the MD (Fig 8),
• Cover the head of the MD with a soft cloth or paper or a wipe.
• Point the head towards the bottom,
• Spray several times by firmly holding the instrument,
• Wipe excess oil with a cloth or a wipe.
VI.4 – Operational test
• Connect the MD to a micro-motor, point the head towards the bottom.
• Turn the micro-motor on at low speed for 30s. Gradually bring up to
maximum speed.
• Wipe excess oil with a cloth or a wipe
x
If you notice overheating, irregularities, vibrations, abnormal noises
when using the MD, immediately contact your AS.
VI.5 - Sterilisation
• The instruments are to be sterilised before first use and after each use.
• Only sterilise cleaned, lubricated and tested instruments.
• We recommend sterilisation that complies with the EN 13060 standard,
class B. All other methods are forbidden.
• Independently put each MD in a sterilisation sachet that complies with
current standards and the size of the MD.
• Comply with the space between the sachets and do not overload the
autoclave.
g
• These MDs and their accessories (apart from accessories with
the logo
minutes minimum (sterilisation time).
x
• Check that the device is completely dry at the end of the cycle.
VI.6 - Storage
D
Keep the MD in a sterilisation sachet in a clean and dry place.
In the event of breakdown, please contact your approved distributor or our
after-sales department directly.
E
Repairs must only be carried out by an approved repairer or by
Anthogyr After Sales Department, only with Anthogyr original
replacement parts.
22
W
which are single use) must be sterilised at 135°C for 18
VII.REPAIR