Integra LifeSciences MAYFIELD A-1047 Mode D'emploi page 10

Kull pins are used in conjunction with all generations of mayfield brand skull clamps (ref a-1013, a-1014, a-1059, a-1108, or a-2000), mayfield brand radiolucent skull clamps (ref a-1058 or a-2002), and the improved gardner brand skull clamp (ref a-1035)
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 15
EN – English
WARNING
• Reusable Skull Pins are sharp. Take necessary precautions when
handling.
Limitations on Reprocessing
• Repeated processing has minimal effects on these devices. Product
life is normally determined by wear and damage due to use.
• It is important to have Integra NeuroSpecialists perform routine
inspections.
INSTRUCTIONS
Containment/Transportation
• Follow healthcare facility protocol for safe containment and trans-
port to the decontamination environment.
• It is recommended that devices are cleaned immediately after use.
Preparation for Cleaning
• See manual wash instructions.
Cleaning
• See manual wash instructions.
Drying
• Products should be dry at this point. If moisture or excess liquid is
detected, dry with a soft clean towel.
• Medically compressed air can be used if needed.
Inspection and Function Testing
• Visually inspect equipment for damage and wear.
Packaging
• Wrapped – using wrappers designed for steam sterilization.
Sterilization Equipment: Steam Sterilizer
Method
1. Follow sterilizer manufacturer's instructions for loading and
operating the sterilizer.
2. Make sure direct steam exposure to all surfaces is possible.
NOTE – Avoid heavy loads and overcrowding on sterilizer shelf/
racks.
3. Steam sterilize at 132.2ºC – 135ºC high vacuum (2psia) for 3
minutes.
4. Remove tray from sterilizer following hospital procedure and
allow cooling.
NOTE – Allow components to reach room temperature before use.
Storage
• Store equipment according to hospital procedures.
NOTE:
Allow skull pins to cool to room temperature before use.
10