Risk of instrument damage due to overheating
An insufficient airflow can cause the R&S RTB2000 to overheat, which may
impair the measurement results, disturb the operation, and even cause
damage.
● Ensure that all fan openings are unobstructed and that the airflow perfo-
rations are unimpeded. The minimum distance to a wall is 10 cm.
● When placing several instruments side by side, keep a minimum dis-
tance of 20 cm between the instruments. Ensure that the instruments do
not draw in the preheated air from their neighbors.
● When mounting the instrument in a rack, observe the instructions of the
rack manufacturer to ensure sufficient airflow and avoid overheating.
2.3

Starting the Instrument

2.3.1

Powering On

The R&S RTB2000 can be used with different AC power voltages and adapts
itself automatically to it.
The nominal ranges are:
● 100 V to 240 V AC at 50 Hz to 60 Hz
● 0.95 A to 0.5 A
● max. 60 W
Risk of injury
Connect the instrument only to an outlet that has a ground contact.
Do not use an isolating transformer to connect the instrument to the AC
power supply.
Getting Started 1333.1605.02 ─ 02
Preparing for Use
Starting the Instrument
11