R&S
RTB2000
1. Connect the power cable to the AC power connector on the rear panel of the
R&S RTB2000.
2. Connect the power cable to the socket outlet.
3. Switch the main power switch at the rear of the instrument to position I.
The STANDBY key lights up. The key is located in the bottom left corner of
the front panel.
You can leave the main power switch on to preserve your last instrument settings.
To disconnect from power supply, power off the instrument.
2.3.2
Starting Up and Shutting Down
To start up the instrument
1. Make sure that the R&S RTB2000 is connected to the AC power supply and
the main power switch on the rear panel is in position I.
2. Press the STANDBY key. The key is located in the bottom left corner of the
front panel.
The instrument performs a system check and starts the firmware. If the previ-
ous session was terminated regularly, the oscilloscope uses the last settings.
Table 2-1: Colors of the STANDBY key
Green
Yellow
Warm-up and prepare the instrument
Make sure that the instrument has been running and warming up before
you start the self-alignment and the measurements. The minimum warm-up
time is about 20 min.
To shut down the instrument to standby state
► Press the STANDBY key.
All current settings are saved, and the software shuts down. Now it is safe to
power off the instrument.
Getting Started 1333.1605.02 ─ 02
Instrument is on: firmware is working
Standby: instrument is off, main power switch is on
Preparing for Use
Starting the Instrument
12