140
Chapter 4 Method Parameters
Fig. 89
After the start of titration, it makes frequently sense to have the sample stirred over a defined period of time, for
instance, to allow for the sample to be dissolved. The waiting time to be observed prior to the first addition of
titration solution can be set using the <Initial waiting time> item. The initial waiting time can be set between 0
and 999 seconds:
Fig. 90
Dynamic control
If dynamic control was selected, one has a selection of 3 different stages: steep, medium and flat. The user-
®
defined dynamic parameters can only be set on the TitroLine
7000. On the first three stages, both the dynamic
parameters and the minimum and maximum step sizes are defaulted:
Fig. 91 a