Generally applicable titration parameters
Depending on the titration mode (dynamic, linear, end-point titration, Dead-stop titration and pH Stat titration), it is
possible to enter a variety of parameters. The following parameters are valid for all automatic titration modes:
Measured value (pH, mV, µA)
Measurement speed
Initial waiting time
Pre-titration
Titration end
But please note that the measurement speed and the titration end differ again as a function of the respective
titration mode.
<Measured value> is the first selection to be made. In the present example, the selection is "pH".
Fig. 84
The selected measured value is displayed for information.
Fig. 85
<Measuring speed> or drift will determine the span of time after which the measured value will be accepted
following a titration step.
Fig. 86
Chapter 4 Method Parameters