Compression Load Cell S-Type PR 6241
Cable length
Cable gauge
Cable bend radius
Cable sheath material
Cable sheath color
4.3

Possible marking of the load cell for the Ex area

Zone
Marking
0 and 1
II 1G Ex ia IIC T6 Ga
Ex ia IIC T6 Ga
20 and 21
II 1D Ex ta IIIC T160 °C Da
Ex ta IIIC T160 °C Da
2
II 3G Ex nA IIC T6 Gc
22
II 3D Ex tc IIIC T85 °C Dc
IS CL I, II, II, DIV 1, GP A, B, C, D, E, F, G Entity -
4012 101 5688
NI CL I, II, II, DIV 2, GP A, B, C, D, E, F, G NIFW -
4012 101 5688
T4A Ta= -30°C to 70°C; T5 Ta= -30°C to 55°C
IS CL I, II, II, DIV 1, GP A, B, C, D, E, F, G Entity -
4012 101 5688
NI CL I, II, II, DIV 2, GP A, B, C, D, E, F, G NIFW -
4012 101 5688
T4A Ta= -30°C to 70°C; T5 Ta= -30°C to 55°C
Minebea Intec
5 m
2
4×0.35 mm
≥25 mm (ixed installation)
≥75 mm (lexible installation)
Thermoplastic elastomer (TPE)
Gray (standard version)
Blue (Ex version)
NOTICE
Installation in the Ex area
For installations in the Ex area, it is imperative to observe the Ex safety instructions
in the installation manuals.
Certiicate no.
for
BVS 16 ATEX E 005
only PR 6241/..E
IECEx BVS 16.0005
TÜV 03 ATEX 2301X
all PR 6241 without
IECEx TUN 17.0025X
/..E
MIN16ATEX001X
all PR 6241 without
/..E
MIN16ATEX001X
all PR 6241 without
/..E
FM17US0276
all PR 6241 without
/..E
FM17CA0138
all PR 6241 without
/..E
4 Speciications
EN-10