Design And Function; Controls And Indicators - Mettler Toledo LV12 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 72

2 Design and Function

The LV12 is an automatic feeder for capsules, tablets or small electronic and mechanical
components. Large weighing series are radically simplified. The LV12 is therefore especially suitable
for statistical quality control (SQC). Usually the LV12 is remotely controlled by commands over a
serial interface (RS232 or USB), but the system can also be controlled manually using push buttons.
The LV12 delivers a single part to the balance and then stops automatically. After receiving the appro-
priate command from the weighing instrument, the device delivers the next part. This process can be
stopped at any time by a command. At the end of a sample, the feeder plate can be emptied or the
feeder waits for the command for a further sample.

2.1 Controls and indicators

Controls
Button
Start/Stop/
Discharge
Feed Rate/Test
Note
By pressing both buttons at the same time, the part size setting can be changed, see [Configure
feeding rate sets } Page 17].
Indicators
LED indicator
Ready
Feed
6
Design and Function
Action
Press once
Press and hold until the level indicating
LEDs flash
Press once
Press and hold until the LED Ready
flashes
State
Green LED lit up
Green LED flashes
Orange LED lit up
Orange LED flashes
One LED lit up
Two LEDs lit up
Three LEDs lit up
Objective
To switch feeding action on/off
To switch on discharge
Change feed rate by one step
Activate the test mode to find the optimal
feed rate
Meaning
Instrument is switched on and ready for
operation
Instrument is in test mode
Feeding action is switched on and the
feeder plate vibrates
Light barrier is interrupted, dirty or faulty
Lowest feed rate
Lower intermediate feed rate, default after
switching on
Upper intermediate feed rate
Automatic Feeder

Publicité

Table des Matières
loading

Table des Matières