7.4. AUTOSTERIL automatic disinfection system
Description of the system.
This system performs an automatic disinfection cycle for the water circuits
of the following instruments:
• all instruments found on the instrument board
• the syringe and instrument placed on the assistant's board,
• the suction tubes (if the S.H.D. system is present),
• water to cup ducts
The system is equipped with a tank [ a ] positioned in the hydro group with
a capacity of about 970 cc of disinfectant liquid of which 230 cc is reserve.
The disinfection cycle can be set and is equipped with an electronic safety
system in compliance with CEE 93/42, and subsequent changes, Medical
Device Directive.
ATTENTION!
• Perform a disinfecting cycle at the end of each work day.
It is strongly unadvisable to let the disinfectant liquid dwell in the
ducts for a contact time of more than 30 minutes.
Low disinfectant liquid warning signal.
When the disinfectant liquid in the tank [ a ] is about to run out, a warning
icon ( D ) appears on the dentist's instrument board console.
NOTE: the icon ( D ) is black if the AUTOSTERIL system is turned
on while it is red if the AUTOSTERIL system is turned off.
Filling the tank containing disinfectant liquid.
To fi ll the tank containing disinfectant liquid operate as follows:
• Remove the tank [ a ]
• Take off the cap [ k ] and pour hydrogen peroxide into the tank until it is full.
NOTE: the shape of the cap allows it to be used as a funnel to
more easily fi ll the tank.
• Put the cap and tank back into place.
• Close the side cover of the hydrogroup.
ATTENTION!
Fill the tank only with pure PEROXY Ag+ or 3% oxygenated water (10
volumes) without diluting.
Setting the disinfecting cycle.
• Check the level of liquid disinfectant in the tank, topping up if necessary.
NOTE: the disinfecting cycle will not start if the liquid in the tank
is below the reserve level.
• Use the doctor's console or button AS found on the assistant's board to
go to the "AUTOSTERIL cycle setup" menu and set the time the liquid
disinfectant has to remain in the water ducts of the instruments (see
paragraph 5.1.1.3.1.).
• Place the instruments to be disinfected in the container provided in the
hydrogroup.
ATTENTION!
Use the adapter [ f ] provided for the syringe. The heater should be
turned off.
The micromotor should be put in without the handpiece.
• In order to disinfect the suction tubes, insert the suction tube terminals in
the fi ttings found underneath the manifold (see paragraph 7.5.).
NOTE: make sure the suction tube terminals are opened.
• If the water to cup ducts need to be disinfected, place the container [ e
] provided under the cup spout.
• Make sure the spray cocks [ d ], found at the bottom of the board, are
opened.