How to remove the separator's container
ATTENTION!
Gloves must be worn when carrying out the following operation to
prevent contact with infected material.
• Move the dental chair fully up.
• Open the cover at the side of the hydrogroup [ a ] (see paragraph 7.7)
and turn the electrical box [ b ].
• Turn the electrical box [ b ] and hydrogen peroxide tank [ a ] (if present).
• Completely empty the separator's container by pressing the timed button
[ c ] found on the cover.
• If present, remove the valve [ s ] for central systems.
• Turn and raise the container until it is detached from the drain pump (k).
• Detach the container [ d ] from the cover [ f ] by pulling up the two side
elastic bands [ e ].
•
After the cleaning operations, refi t the vessel [ d ] after lubricating the
O-rings with S1-Protective Lubricant for O-Rings (CEFLA S.C.).
• Put the electrical box [ b ] back in place and put the cover [ a ] back on
the side of the hydrogroup (see paragraph 7.7).
9.6. Cleaning the turbine return air fi lter
Monthly check the oil container fi lter [ g ] present in the turbine's return
air line.
If necessary, replace the fi lter element (code 97290014).