CASTELLINI SKEMA 6 CART Instructions De Service page 108

Tablette praticien
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Operation with the instruments withdrawn
• Lever [2] to the right: starts the instrument and adjusts its speed/power.
NOTE: if the camera is withdrawn, the image is frozen. .
• Lever [2] all the way to the left: CHIP-AIR for the turbine or the micromotor
is started.
The air jet is stopped when the lever [2] is released.
NOTE: if the camera is withdrawn, the saved images can be
selected one after the other.
• Hold down the pedal [4] to start the CHIP-WATER for the turbine, micro-
motor or scaler.
The water jet is stopped when the pedal [4] is released.
NOTE: if the camera is withdrawn, full-screen display of the se-
lected image is activated/deactivated.
Water spray to instruments on/off
The spray is delivered by holding down the pedal [4] when the instru-
ment is turned on (lever [2] activated).
NOTE: the type of spray delivered can be set from the doctor's
console.
NOTE: you can set the foot control in such a way that the control
[4] sets instrument spray activation or deactivation before you operate the
lever [2] (bistable operation).
Only a CASTELLINI authorized technician can set the foot control in this
manner.
NOTE: at the end of work a blast of air is automatically activated to
eliminate any residual drops of liquid in the instrument ducts.
Joystick for dental chair movements [3]
The joystick for dental chair movements [3] has 2 different operating modes,
depending on whether or not the pedal [4] is pushed before hand.
NOTE: once the pedal [4] is pushed, the joystick [3] must be ac-
tuated within 5 seconds. The function is automatically cancelled once this
time has elapsed.
Operation with the instruments in place on the holder
The joystick [3] actuates the 4 manual dental chair movements.
NOTE: release the joystick to stop manual movement.
Operation with the instruments in place on the holder after pushing the
pedal [4]
The joystick [3] retrieves the 4 dental chair programs.
NOTE: actuate the joystick again to stop automatic movement.
With instrument withdrawn but not turned on
The joystick [3] actuates the 4 manual dental chair movements.
NOTE: the 4 programs are blocked under this condition.
With instrument turned on (lever [2] actuated)
Joystick disabled, all the commands to move the dental chair are blocked.
Foot control setup.
The foot control has 4 different confi gurations identifi ed as 0, 1, 2 and 3:
0 = Joystick for only direction operation of the dental chair
1 = Auxiliary equipment and dental chair controllers (default confi guration)
2 = ENDO mode and micromotor reverse rotation
3 = Surgical mode
Operation in the various confi gurations is explained below.
NOTE: the confi guration of the foot control can be changed only by
a CASTELLINI authorized technician.
108
108
108
GB
EN
SKEMA 6
- OPERATING INSTRUCTION

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Skema 6 cpSkema 6 rs

Table des Matières