Resetting the calibration
1. To recalibrate the device, you need the SensoLite 310. Mount the
SensoLite at the level of reference point A2 (see previous page). Carry
out the calibration with the Commander DAL 50.
Important: Use the buttons on the calibration panel.
2. During calibration, pay attention to the alignment of the axes on the
DualAxis laser. Always calibrate all the axes.
3. Switch the device to calibration mode:
Switch off the DualAxis laser and then switch it on again by pressing the
auto/man. button (32) (press ON/OFF button (34) briefly). Keep the
auto/man. button (32) pressed until the message "CAL" (36) appears on the LCD
display. Then release the auto/man. button (32) again. You can toggle between the
axes with the X/Y button (38) on the Commander DAL 50. The laser begins to rotate
as soon as the X/Y button is pressed.
4. Correcting the calibration:
With the plus/minus buttons (39) on the Commander DAL 50, you can move the laser
away from its current position at the level of reference point A2.
5. Ending the calibration:
Cancel: The entire calibration is aborted and the previous status is restored if the
device is switched off before the Enter button has been pressed.
Save: The new calibration is saved by pressing the Enter button (40) on the
Commander DAL 50. After saving, switch off the DualAxis laser. The new setting
becomes active when the device is switched on again.
TIP: Save every axis individually.
Note:
i
Regularly check the calibration
before use, after transport and after
extended periods of storage. Always
make sure to control all axes.
Y
ENGLISH
DualAxis-Laser
X- / Y- axis
X