3-4 Ω RESISTANCE MEASUREMENT
Zero adjustment on the ohmmeter is carried out using the calibration button (front) by
short-circuiting the inputs.
Ω
Scale (indication)
Reading coefficient
Measuring scale
Internal resistance
End of scale current
Open circuit voltage
Accuracy
Admissible overload
(1) Buzzer for R ≤ 100, no buzzer for R ≥ 250Ω
(2) of the mid-scale value
(3) for 5 seconds (protection by Resistances and CTP)
(4) for battery voltage of 1.45V ± 0.1V

3-4-1 AUDIBLE CONTINUITY TEST - SEMI-CONDUCTOR TEST

NB : In Ω x 1: "buzzer" threshold ≤ 250Ω, control of diode on or blocked (
in "COM" for on). Do not carry out live measurements.
On the ohmmeter, calibre Ω x 1, if voltage of V
issues a modulated sound until the error has been corrected.
4-1 Dimensions and weight
97 x 155 x 43 mm
420 g
4-2 Power supply
1.5V battery (R6 saline or LR6 alkaline)
Battery life:
about 250 hours for the ohmmeter with an alkaline battery
4000 5-second measurements for R < 50Ω (with beep)
4-3 Maximum climatic conditions
Temperature:
Relative humidity:
Altitude:
Ω x 1 (1)
x 1
10Ω to 20kΩ
200Ω
8mA

4 - GENERAL CHARACTERISTICS

use -10°C to +50°C ; storage -30°C to +70°C
use ≤ 80% HR
use < 2000m
Ω x 10
20k...0 (c)
x 10
100Ω to 200kΩ
2kΩ
800µA
1.5V
±10% (2) (4)
400V (3)
> 5V (rms) is applied, the "buzzer"
AC
15
Ω x 100
x 100
1kΩ to 2MΩ
20kΩ
80µA
, anode