Resistance
measurement Ω Ω Ω Ω
Audible continuity
test - semi-conductor
test
General characteristics
Dimensions and
weight
Power supply
Maximum climatic
conditions
Compliance with
international
standards
Electromagnetic
compatibility
18
Zero adjustment on the ohmmeter is carried out using the calibration
button (front) by short-circuiting the inputs.
Ω Ω Ω Ω
Scale (indication)
Reading coefficient
Measuring range
Internal resistance
End of scale current
Open circuit voltage
Accuracy
Admissible overload
(1) Buzzer for R ≤ 100, no buzzer for R ≥ 250Ω
(2) Of the mid scale value
(3) For 5 seconds (protection by resistors and CTP)
(4) For battery voltage of 1,45V ±0.1V
NB : In Ω x 1 : "buzzer" threshold ≤ 250Ω, control of diode on or blocked
(
, anode in "COM" for on). Do not carry out live measurements.
On the ohmmeter, calibre Ω x 1, if voltage of VAC > 5V (rms) is applied,
the « buzzer » issues a modulated sound until the error has been
corrected.
•
97 x 155 x 43 mm
•
420 g
•
1.5V battery (R6 saline or LR6 alkaline)
•
Battery life:
about 250 hours for the ohmmeter with an alkaline battery
4000 5-second measurements for R < 50Ω (with beep)
•
Temperature
•
Relative humidity
•
Altitude
Electrical safety (EN 61010-1 - 2001)
CEI 1010-1 EN61010 NF-C 42020 VDE 0411
•
Double insulation :
•
Pollution level :
•
Installation category : III according to CEI 664
•
Allocated voltage : 600V
•
Emission (EN 61326-A2 – 2001)
•
Immunity (EN 61326-A2 – 2001)
Maximum influence in the presence of conducted radio frequencies:
3 times the accuracy class if the length of the measured circuit is >
3m.
Ω x 1 (1)
Ω x 10
20k...0 (c)
x 1
X 10
10Ω to 20kΩ
100Ω to 200kΩ
200Ω
2kΩ
8mA
800µA
1.5V
±10% (2) (4)
400V (3)
use -10° C to +50° C ; storage -30° C to +70° C
use ≤ 80% HR
use < 2000m
2
English
Ω x 100
x 100
1kΩ to 2MΩ
20kΩ
80µA
Multimeter MX1