Siemens SITRANS F FUT1010 Instructions De Service page 43

Débitmètres à ultrasons. débitmètres de gaz ip65 nema 4x, ip66 nema 7
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Error or Message
Low Signal - Press <ENTER>
Detection Fault
Note
If you receive a Detection Fault message, it is strongly recommended that the Technical
Service Department (http://www.automation.siemens.com/partner) be contacted.
FUT1010 IP65 NEMA 4X & IP66 NEMA 7 Gas Quick Start
Operating Instructions, 1/2013, A5E02639183-AC
Probable Cause
During the Initial Makeup the flow
meter decides that the level of the
receive signal is insufficient for proper
operation.
If it appears that the flow meter cannot
complete an Initial Makeup it means
that the pipe and/or gas conditions do
not permit a receive signal that meets
the flow detection standards.
The system will not operate.
Troubleshooting
5.1 Troubleshooting
Solution
Some reasons for low signal are:
Invoking [Install Complete] on an
empty pipe.
Coupling compound insufficient; not
applied or evaporated. Reapply
couplant.
A disconnected or broken sensor
cable.
The pipe needs to be conditioned at
the mounting location.
Insure that line is sufficiently
pressurized.
The sensor cables are defective or
not connected to the correct
channel.
If you locate and correct the improper
condition immediately, press <ENTER>
to resume the installation procedure.
Otherwise, press the <Left Arrow> to
abort the installation and conduct a
thorough investigation.
Attempt to improve operating
conditions by reinstalling the sensors at
a different spacing offset, or even at a
different location on the pipe.
41

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Sitrans f 7me363

Table des Matières