9. Fuse Replacement
If the Control Unit does not function, check the fuses (Fuse Box lock
located on the rear of the Control Unit). To access the Fuse, use
a pointed tool push on the fuse locking latch and the drawer will
spring open. (Fig.16)
10. Maintenance
10-1 Maintenance of the Control Unit and Foot Control
If blood or saline solution has stained the Control Unit or Foot Control, remove the AC Electrical Cord; wipe Control
Unit or Foot Control with a damp cloth, then with an alcohol-absorbed cloth.
10-2 Maintenance of the Handpiece and Motor
This handpiece and motor can be cleaned and disinfected with
a Thermo-Disinfector.
Attach the Protection Plug to the motor.
For detail, refer to the Thermo-Disinfector's Operation Manual.
CAUTION
• Always use the Protection Plug during Thermo-Disinfector.
• Do not use Protection Plug without the mounting O-Ring. It may cause a malfunction. If O-Ring had damaged,
replace it immediately.
NOTICE
Protection Plug O-ring is consumable supply. If it damaged, replace it new one.
Optional Protection Plug: Order Code 10001595
Protection Plug O-Ring: Order Code 0312457102
Protection Plug
Protection Plug O-Ring
18
Fuse
Fig.16
Fig.17