Sterilization - NSK Surgic Pro Manuel D'utilisation

Masquer les pouces Voir aussi pour Surgic Pro:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

11. Sterilization

This handpiece and motor can be autoclavable up to Max.135°C.
• Autoclave sterilization is recommended.
• Autoclave sterilization is required for the first time use and after each patient as noted below.
CAUTION
Only Items Specified below can be autoclaved.
The following items can be autoclaved.
• Handpiece
• Internal Irrigation Nozzle
• Protection Plug
*Autoclaving
1)
Remove blood and debris from the handpiece and motor.
2)
Clean inside the handpiece; by using the spray lubricant
3)
Place those in autoclave pouch (not included in the package) and seal it.
4)
Autoclavable up to max. 135˚C.
ex.) Autoclave for 20 min. at 121°C, or 3 min. at 134°C.
5)
Keep the handpiece in the autoclave pouch to keep it clean until you use it.
* EN13060/EN ISO17665-1 recommends autoclaving for 3 minutes (minimum holding time) at 134°C or 15
minutes(minimum holding time) at 121°C.
NSK recommends Class B sterilization.
CAUTION
• Clean and lubricate the handpiece before autoclaving. Autoclaving a handpiece stained with blood or debris could
cause damage to the handpiece.
• Do not lubricate the micromotor.
• Do not disconnect the motor cord from the micromotor.
• The Irrigation tube is a single use and cannot be autoclaved.
• We do not recommend you to sterilize on plasma sterilization or EOG sterilization.
• Micromotor with Motor Cord
• Tube Holder
• Calibration Bur
19
• Handpiece Stand
• Nozzle Holder

Publicité

Table des Matières
loading

Table des Matières