Factory Settings - NSK Surgic Pro Manuel D'utilisation

Masquer les pouces Voir aussi pour Surgic Pro:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

NOTICE
• Set the Clock before you use this system.
• If the clock display does not appear or is shown as "- - : - -" the internal batteries may have run out. Contact the
retailer where device was purchased.
• The clock runs on a 24-hour display system.
• The years can be set from 2010 to 2099.
1)
Push both the USB REC key and SPEED (-) for about 2 seconds.
2)
The setting screen will appear on the LCD. Settings can be made for year, month, day, hour, and minute in that
order. Input the data, and press the USB REC key to move to the next item. (The item you have chosen will blink.)
3)
Change numbers by pressing the SPEED (+ -) key. A long push will increase numbers in increments of 5.
4)
When all items are set, press the USB REC key for 2 seconds to save settings.
CAUTION
• If the power goes off while settings are being made, the changes will not be saved. Start once more at the beginning.
• If data is already stored in the internal memory, it will not be possible to set the date.
5-2 Illumination Adjustment (X-SG20L Only)
Repeat press LIGHT key to cycle through light levels. Pressing LIGHT key to change illumination volume. (OFF
LOW
HIGH) (Default is HIGH)
To save the changes, press the MEMORY key for at least 2 seconds.
NOTICE
• Repeat the above procedures when restoring operation.
• The lamp status will remain set even if the power is switched "OFF" and "ON".
• This function is only applicable for an optic motor and handpiece.
5-3 Sound Volume
It is possible to change the volume for operation and warning sounds for keys.
1)
Turn on the power while pressing the [CAL] key.
2)
When turning ON, the volume will alternate between "Large" and "Low." (The default is "Low")
* Sound Volume is not displayed on LCD. The actual sound changes.
NOTICE
Settings will be saved even if the power is turned off.

5-4 Factory Settings

When initialized, the device will revert to factory settings. Preferred settings and programs, etc. will be erased, so
be sure to make memos of them.
1)
Turn power on while pressing the PROGRAM (+ -) Key.
2)
"Set" will come up on the LCD Display. (there will also be a short beep) By pressing [CAL] key, the device will
revert to factory settings.
3)
"Fin" will come up on the LCD Display. (there will also be a long beep) Initializing is complete when the screen
goes back to its usual display.
CAUTION
Even when the device reverts to factory settings, date and time data will not be initialized.
11

Publicité

Table des Matières
loading

Table des Matières