CAUTION
During a process, it is important to hold the Calibration Bur vertically without applying any forcible power. Making
the Calibration Bur tilt or pressing it may cause a loss in accuracy.
Calibration fail example
• Extreme gear abrasion and mechanical loss can exceed the control limits. Correct data cannot be obtained during
calibration cycle and the calibration will fail. In this case, request repair.
• If nonconformities such as extreme abrasion of the gear and mechanical loss exceed the control limit, correct
data cannot be obtained during calibration will fail. In this case, request repair.
• Calibration fail; if the calibration bur makes contact with something during no load calibration or no load is
erroneously applied during load calibration.

6-4 Standard Operation

1)
Select the desired program number.
Control Unit: Press PROGRAM (+ -) Key.
Foot Control: Step on the Foot Control PRG (Program) Pressing the PRG (Program) Button for 1 second
more will descend the to the next program number.
2)
Step on the Speed Control Pedal in the middle of the Foot Control the motor will start to run. The Coolant Pump
will also run (if programmed). Speed increases as the pedal is depressed. When the Speed Control Pedal is fully
depressed the speed reaches the maximum preset value.
3)
Reversing the motor rotational direction: To reverse direction of the motor (and bur) simply step on the Foot
Control Forward/Reverse Button. A warning beep can be heard when the rotational direction is in reverse mode.
CAUTION
• Using the Micromotor with light 'ON' for more than 3 minutes continuously, motor will become hot (The maximum
surface temperature might reach 45˚C by the situation used). In such cases, stop using the Micromotor till the
light part cools down, or use the Micromotor without turning on the light.
• Activation of the torque limiter: During use, when the drilling load reaches the preset torque upper limit, the integrated
torque limiter automatically activates to prevent over torque. When the torque limiter activates, the motor stops after
beeping for 1 second. To reactivate the Micromotor, release the speed control pedal and depress it again.
4) Release the Foot Control Pedal, and the Micromotor will automatically stop.
6-5 USB Function (Surgic Pro + ONLY)
NOTICE
When the clock is not set, the time and date will not be recorded on data. It is recommended that the clock be set
before using the device.
The internal clock is used along with the different operations of this device (date used, starting time, length of time,
Gear Ratio,level of Coolant Solution Flow, Rotation Speed, FWD/REV, Torque, Program Number), making it possible
to store data in chronological order on USB flash drive (CSV files, BMP files).
Stored data can be viewed on a personal computer.
*Files and graphs saved
Folders are created in
order by date
CSV file
CSV file