8-4 Factory Settings

When initialized, the device will revert to factory settings. Preferred settings and programs, etc. will be erased, so be
sure to make memos of them.
1) Turn power on while pressing the PROGRAM (+ -) key.
2) "SEt" will come up on the LCD Display. (there will also be a short beep) By pressing BURST key, the device will
revert to factory settings.
3) "FIn" will come up on the LCD Display. (there will also be a long beep) Initializing is complete when the screen
goes back to its usual display.
9

Maintenance

After each patient maintain the product as follows.
9-1 Auto Cleaning
1) Remove the Irrigation Needle from the solution pack.
2) Place the Irrigation Needle in distilled or deionized water in an open container.
3) Place the Tip of the Handpiece into the water.
4) Start Auto-Cleaning by holding down the AUTO CLEANING key for
approximately one second.
While Auto-Cleaning is active "CLn" appears in the POWER Display
on the LCD Display. The Bar Graph indicates the time remaining (Auto
Cleaning lasts for approximately 30 seconds).
*Auto Cleaning can be stopped anytime by pressing AUTO
CLEANING key.
5) Once Auto Cleaning has completed, the LCD Display returns to
normal display.
Perform Auto Cleaning after every use. If Auto Cleaning is omitted, the Handpiece may clog or malfunction.
CAUTION
9-2 Cleaning the LED handpiece with cord
1) Detach the Handpiece Cord from the Control Unit.
2) Remove the Irrigation Tube from the Handpiece and dispose.
3) Remove any debris from the product. Do not use a wire brush.
4) Wipe clean with alcohol.
This icon denotes that the product can be washed via Thermo Disinfector.
At the print date of this Operators Manual, the Miele (Model:G7882) is the only Thermo-Disinfector verified for use with
the VarioSurg 3.
Refer to the Thermo-Disinfector manual.
When using in a Thermo-Disinfector, ensure that the Handpiece and Handpiece Cord are stable by placing
them in a basket or container.
CAUTION
Never use solvents, such as benzene and thinner for cleaning.
The Handpiece can not be removed from the Handpiece Cord.