Linearity Calibration; Geo Adjustment - Uline Ohaus Defender 5000 Mode D'emploi

Masquer les pouces Voir aussi pour Ohaus Defender 5000:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

4. Press the softkey
corresponding to the
icon to initiate
Span Calibration.
(See Figure 10)
5. Place a calibration
mass of the specified
weight on pan and
press the softkey
corresponding to the
icon. To change to a
different calibration point, input desired value
and place the corresponding weight on pan for
calibration. A suggestive message shows on the
screen. (See Figure 11)
6. The message
"Completed" will
display on screen.
(See Figure 12)
7.
Exit Span Calibration
by pressing the softkey
corresponding to the
icon. (See Figure 13)
8. To return to the Main Menu, press the softkey
corresponding to the

LINEARITY CALIBRATION

Linearity calibration uses three calibration points. The
full calibration point is established with a weight on
the scale. The mid calibration point is established with
a weight equal to half of the full calibration weight on
the scale. The zero calibration point is established with
no weight on the scale. The full calibration and mid
calibration points can be altered by the user during the
calibration procedure.
1.
Long press the
User
Menu
button to enter
Main Menu.
(See Figure 14)
2. Press the softkey
corresponding to the
icon to enter the
Calibration submenu.
Scroll to Linearity
Calibration using the
softkey corresponding to
the ▼ icon. (See Figure 15)
PAGE 5 OF 101
OPERATION CONTINUED
Figure 10
Figure 11
Figure 12
Figure 13
icon.
Figure 14
Figure 15
3. Press the softkey
corresponding to
the
icon to initiate
Linearity Calibration.
(See Figure 16)
4. Clear the pan and press the softkey corresponding
to the
icon.
5. Put the calibration
mass of the specified
weight on the pan, and
then press the softkey
corresponding to the
icon for confirmation.
To change to a different calibration point, input the
desired value, and then place the corresponding
weight on pan for calibration. (See Figure 17)
6. Put calibration mass
on pan, and then
press the softkey
corresponding
to the
icon for
confirmation. To change
to a different calibration point, input desired value,
then place the corresponding weight on pan for
calibration. (See Figure 18)
7. The message
"Completed" will
display on screen.
(See Figure 19)
8. Exit Linearity Calibration
by pressing the softkey
corresponding to the
icon. (See Figure 20)
9. To return to the Main Menu, press the softkey
corresponding to the

GEO ADJUSTMENT

Set the GEO factor that corresponds to your location.
GEO codes are numbered 0-31.
1.
Long press the
User
Menu
button to enter Main
Menu. Select the menu
item Calibration by
pressing the softkey
corresponding to the
icon. (See Figure 21)
Figure 16
Figure 17
Figure 18
Figure 19
Figure 20
icon.
Figure 21
0621 IH-8105