Calibration Menu; Span Calibration - Uline OHAUS DEFENDER 3000 Mode D'emploi

Balance numérique
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

RESET
Set the unit menu to the factory default
settings:
NO = Not Reset
YES = Reset
NOTE: If the Legal for Trade
menu item is set ON, the
settings are not reset.
KILOGRAM UNIT
Set the status:
OFF = Disabled
ON = Enabled
POUND UNIT
Set the status:
OFF = Disabled
ON = Enabled
GRAM UNIT
Set the status:
OFF = Disabled
ON = Enabled
OUNCE UNIT
Set the status:
OFF = Disabled
ON = Enabled
PAGE 3 OF 24
OPERATION CONTINUED
POUND OUNCE UNIT
rESEt
Set the status:
OFF = Disabled
NO
ON = Enabled
YES

CALIBRATION MENU

UNIt
kg
Two calibration processes are available – span
calibration and linearity calibration. (See Figure 3)
OFF
Figure 3
ON
lb
UNIt
OFF
1.
ON
2. Make sure that the scale base is level and stable
UNIt
3. Calibration is unavailable with LFT set to ON.
g
4. Allow the Indicator to warm up for approximately
OFF
5. To abort calibration, press the EXIT button anytime
ON
CALIBRATION VALUES
MODEL
UNIt
oz
H-2587
H-4594
OFF

SPAN CALIBRATION

ON
1.
Span
Perform
Linearity
Perform
Geographic
Set 00... Set 19... Set 31
Adjustment
End Calibration Exit CALIBRATE menu
Make sure that appropriate calibration masses are
available before beginning calibration.
during the entire calibration process.
5 minutes after stabilizing to room temperature.
during the calibration process.
SPAN CALIBRATION
#
POINTS (kg.)
5, 10, 15, 20, 25, 30, 40, 50, 60
5, 10, 15, 20, 25, 30, 40, 50,
60, 75, 100, 120, 150
Span calibration uses two points to
adjust the scale. The first point is the
zero value where there is no weight
on the scale. The second point is the span value
where a calibration mass is placed on the scale.
lb
UNIt
oz
OFF
ON
LINEARITY CALIBRATION
POINTS (kg.)
30, 60
75, 150
SPAN
0521 IH-2587