When the indicator is used in trade or a legally
controlled application, it must be set up, verified and
sealed in accordance with local weights and measures
regulations. It is the responsibility of the purchaser to
ensure that all pertinent legal requirements are met.
SETTINGS
Before verification and sealing, perform the following
steps in order:
1. Verify that the menu settings meet the local weights
and measures regulations.
2. Perform a calibration as explained on page 4.
3. Turn off the indicator.
4. Disconnect power from the scale and remove
security screw. (See Figures 96-97)
5. Set the position of the security switch SW1 to on.
6. Close the security switch.
7.
Reconnect power and turn on the indicator.
VERIFICATION
A local weights and measures official or authorized
service agent must perform the verification procedure.
• Verify the input voltage range printed on the data
label matches the local AC power to be used.
• Make sure the power cord does not pose a potential
obstacle or tripping hazard.
• Disconnect scale from power supply when cleaning.
• Do not operate scale in hazardous or unstable
environments.
• Do not immerse scale in water or other liquids.
PAGE 32 OF 101
LEGAL FOR TRADE
SEALING
The local weights and measures official or authorized
service agent must apply a security seal to prevent
tampering with the settings. Refer to the illustrations
below for sealing methods. (See Figures 96-97)
Figure 96
Figure 97
MAINTENANCE
CLEANING
Clean the scale at regular intervals.
Housing surfaces may be cleaned with a lint-free cloth
slightly dampened with water or a mild cleaning agent.
CAUTION! Disconnect the scale from the power
supply before cleaning. Make sure that no
liquid enters the interior of the terminal or base.
NOTE: Do not use solvents, harsh chemicals,
alcohol, ammonia or abrasive cleaning agents
to clean the housing or control panel.
Wire Seal
Security
Screw
Security
Screw
Paper
Seal