The Percent Configurations are defined below.
NOTE: Defaults settings are in bold.
CONFIGURE ITEM
OPTION
Auto Tare
Off/On
Accumulate
Off/Manual/Automatic
INPUT/OUTPUT (I/O) SETUP
The I/O's setup can be customized for user preferences.
ITEM
AVAILABLE SETTINGS
Enable
On, Off
NOTE: I/O's will only work when the I/O Option
board has been installed. The option I/O
board provides two isolated inputs and four
dry-contact normally open relay outputs which
can be used for simple process weighing.
1.
Press the
button
User
Menu
to enter Main Menu.
With the softkey
corresponding to the
▼ icon, go down list
and highlight Application
Mode. (See Figure 73)
2. Enter this submenu by pressing the button
corresponding to the
3. In the Application
Mode menu, enter
the Percent submenu.
(See Figure 74)
4. Select list item and
press the softkey
corresponding to the
icon to change
setting. (See Figure 75)
5. After completion of I/O's setup, press the softkey
corresponding to the
application screen.
PAGE 20 OF 101
APPLICATIONS CONTINUED
DESCRIPTION
Off: Auto tare is turned off.
On: The first stable weight (>=5d) will be tared as a container automatically.
Off: The icon "Σ" does not display.
Manual: The softkey "Σ" displays. User can press the key to accumulate the stable weight.
Automatic: The icon "Σ" displays. The weight will be accumulated automatically.
When LFT is ON (no such limit when LFT is OFF or the approved mode is OIML),
a. Gross weight and net weight cannot be accumulated;
b. Gross 0 has to be reached before a new sample accumulation.
When LFT is OFF,
a. Gross weight and net weight can be accumulated at the same time;
b. A new sample can be accumulated after reaching gross 0 or net 0.
Figure 73
icon. (See Figure 73)
Figure 74
Figure 75
icon to return to the main
NOTE: The load to be accumulated has to be >= 10d. The next
accumulation can only start once the pan is cleared.
DYNAMIC WEIGHING
Use this application to weigh an unstable load, such as
a moving animal. Three different start/reset operation
types can be selected: Manual (start and stop via key
press), Semi-Automatic (auto-start with manual reset),
and Automatic (start and stop automatically).
1.
Press the
Library
Dynamic is displayed in the upper left corner of the
screen. (See Figure 76)
2. Press the softkey
corresponding to
the
icon to start
averaging. To abort
the averaging, press the
button corresponding to the
3. When the averaging has finished, press the button
corresponding to the
APPLICATION SETUP
The Application can be customized for user preferences.
1.
Press the button
corresponding to the
icon to enter
Configuration Menu.
(See Figure 77)
2. Select the list item
and press the softkey
corresponding to the
icon to change
setting. (See FIgure 78)
3. To return to the Application home screen, press the
softkey corresponding to the
button until the icon corresponding to
Mode
icon. (See Figure 76)
icon to reset.
icon.
Figure 76
Figure 77
Figure 78
0621 IH-8105