The Dynamic Configurations are defined below.
NOTE: Defaults settings are in bold.
CONFIGURE ITEM
OPTION
Dynamic Mode
Countdown/Continuous
Operation Type
Manual/
Semi-Automatic/
Automatic
Start Weight
3,000 (with current unit)
Auto Tare
Off/On
Accumulate
Off/Manual/Automatic
Duration Time
1 ~ 10 s
PAGE 21 OF 101
APPLICATIONS CONTINUED
DESCRIPTION
Countdown: There is a countdown time.
Continuous: There is no countdown time.
Manual:
Place load on the pan.
Press softkey
to start dynamic weighing and countdown. When countdown is over,
average weight is displayed. Before doing next weighing, press softkey "Reset".
Semi-Automatic:
Before weighing, the reading should be zero (Gross or Net). Place load (>=Start Weight) on
the pan. The dynamic weighing and countdown start automatically. When countdown
is over, average weight is displayed. Before doing next weighing, press softkey "Reset".
Automatic:
Before weighing, the reading should be zero (Gross or Net). Place load (>=Start Weight)
on the pan, the dynamic weighing and countdown start automatically. When the
countdown is over, average weight is displayed. After the load has been removed,
the average weight still displays until the duration time is over. If the softkey "Reset" is
pressed, average weight will be cleared immediately.
NOTE: During the weighing countdown, if a weight error
(underload/overload) has occurred, the weighing process will
stop immediately. If Auto Tare is on, a container (weight >=5d)
must be placed on the pan first. After the terminal has done taring
automatically, place load on the pan to start dynamic weighing.
If the operation type is Semi-Automatic/Automatic, this menu will be shown.
If the load is bigger than the weight, then the dynamic weighing will start.
Off: Auto tare is turned off.
On: The first stable weight (>=5d) will be tared as the container weight.
Off: The softkey "Σ" doesn't display.
Manual: The softkey "Σ" displays. Press the key to accumulate the stable weight.
Automatic: The softkey "Σ" displays. Weight will be accumulated automatically.
NOTE:
1. The load to be accumulated has to be >= 10d. Another
accumulation can't be done until the pan is cleared (< 5d).
2. Gross weight and net weight can't be accumulated together
when the LFT is ON (no such limit when the LFT is OFF or the
approved mode is OIML). When first accumulated value is
gross weight (net weight), scale will enter into gross weight
(net weight) accumulate mode. Otherwise, "Gross and net
weight can't be accumulated" will be displayed.
3. If LFT is ON, gross weight 0 has to be reached before a new
sample can be accumulated. If LFT is OFF, a new sample
can be accumulated after reaching gross or net weight of 0.
It is the time for the display to remain the dynamic weighing result after
load is removed.