SERIAL COMMUNICATION AND PRINTER SETUP CONTINUED
HANDSHAKE
Set the flow control method.
NONE
= no handshaking
XON/XOFF = XON/XOFF handshaking
HARDWARE = hardware handshaking
(COM1 menu only)
ALTERNATE PRINT COMMAND
Set the alternate command character for Print.
Settings of A(a) to Z(z) are available.
The default setting is P.
ALTERNATE TARE COMMAND
Set the alternate command character for Tare.
Settings of A(a) to Z(z) are available.
The default setting is T.
ALTERNATE ZERO COMMAND
Set the alternate command character for Zero.
Settings of A(a) to Z(z) are available.
The default setting is Z.
RESET
Reset the settings to factory default.
PRINT SETUP OF RS-232
DEMAND
If Demand is selected, the submenu Stable Only will
display.
Set the printing criteria.
OFF = Values are printed immediately, regardless
of stability.
ON = Values are printed only when the stability
criteria are met.
AUTO ON STABLE
If Auto On Stable is selected, the submenu Mode will
display.
Set the printing mode.
Load = Prints when the displayed load is stable.
Load and Zero = Prints when the displayed load
PAGE 24 OF 101
and zero reading is stable.
AUTO ON ACCEPT
If Auto On Accept is selected and weighing mode is
Check, values will be printed when weight is accepted.
ACCEPT = Printing occurs each time the display is
within the Checkweigh accept range
and stability criteria are met.
INTERVAL
If Interval is selected, the submenu Time will display.
INTERVAL = Printing occurs at defined time interval.
The time interval can be set through the numeric
keypad. Settings of 1 to 3,600 seconds are available.
Default is 1. Printing occurs at the defined time interval.
MT-CONTINUOUS
If MT-Continuous is selected, the print output will be in
the MT-Continuous format.
CONTINUOUS = Printing occurs continuously.
NOTE: Refer to Appendix A for MT-Continuous
format.
Checksum
Off = disabled
On = enabled
OH-CONTINUOUS
If OH-Continuous is selected, the print output will be in
the OH-Continuous format.
NOTE: Refer to Appendix A for OH-Continuous
format.
CONTINUOUS = Printing occurs continuously.
SICS
OFF = disable MT-SICS command
ON = enable MT-SICS command
NOTE: Refer to Appendix B for SICS commands.
REFERENCE BALANCE
OFF = Do not connect to reference balance.
ON = Connect to reference balance.
NOTE: Use a reference balance to perform
sampling with a high resolution balance in
Counting Mode. Make sure the balance is already
on before connecting to the TD52 indicator.