AVERAGE TIME SETUP

There are two methods to set the averaging time. The
default average time is 10 seconds.
NOTE: When time is set to 0, the first stable
weight over 5d is displayed. Averaging time
can be set to between 0 and 20 seconds.
Method 1
1.
Press the button
corresponding to
the
icon to enter
Dynamic submenu to
change averaging time.
(See Figure 79)
2. Press the softkey
corresponding to the
icon. Then enter
desired value using
keypad. (See Figure 80)
3. After setting the
desired averaging
value, press the
softkey corresponding
to the
icon for
confirmation. (See Figure 81)
4. Press the softkey
corresponding to the
icon to exit
submenu. (See Figure 82)
Method 2
1.
Use the alphanumeric
keypad to key in the
desired averaging
time. (See Figure 83)
2. Once the value
keyed in is displayed
in upper left portion of
the screen, press the
softkey corresponding
to the
icon.
(See Figure 84)
PAGE 22 OF 101
APPLICATIONS CONTINUED
INPUT/OUTPUT (I/O) SETUP
The I/O's setup can be customized for user preferences.
The I/O's setup is defined below.
Figure 79
Figure 80
1.
Figure 81
2. With the button corresponding to the ▼ icon,
3. To enter this submenu, press the softkey
Figure 82
4. In the Application
5. Select the list item
Figure 83
6. After completion of I/O Setup, press the softkey
Figure 84
NOTE: Defaults settings are in bold.
ITEM
AVAILABLE SETTINGS
Enable
On, Off
Discrete Input 1
Off, Zero, Tare, Clear Tare, Print, Start, Reset
Discrete Input 2
Off, Zero, Tare, Clear Tare, Print, Start, Reset
Discrete Output 1
Off, Overload, Underload
Discrete Output 2
Off, Overload, Underload
Discrete Output 3
Off, Overload, Underload
Discrete Output 4
Off, Overload, Underload
NOTE: The I/O's setup will only work when the I/O
Option board has been installed. The option
I/O board provides two isolated inputs and four
dry-contact normally open relay outputs which
can be used for simple process weighing.
User
Press the
button
Menu
to enter the Main
Menu. (See Figure 85)
go down the list and highlight Application Mode.
corresponding to the
Mode menu, enter the
Dynamic submenu.
(See Figure 86)
and press the softkey
corresponding to the
icon to change the
setting as desired.
(See Figure 87)
corresponding to the
application screen.
Figure 85
icon.
Figure 86
Figure 87
icon to return to the main
0621 IH-8105

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

H-8105H-8106H-8107