OPERATING ISSUE
EEP Error
Unit will not
turn on.
Cannot zero the
scale, or scale
will not zero
when turned on.
Unable to
calibrate.
Cannot display
weight in desired
weighing unit.
Cannot change
menu settings.
Error 8.1
Error 8.2
Error 8.3
Error 8.4
Error 8.6
Error 9.5
Battery symbol
flashing
CAL E
NO.SW
REF WT Err
If the troubleshooting section does not resolve your problem, contact ULINE Customer Service at 1-800-295-5510.
PAGE 33 OF 101
TROUBLESHOOTING
CAUSES
EEPROM checksum error
Power cord not plugged in or connected.
Power outlet not supplying electricity.
Battery discharged (T52P).
Other failure.
Load on Scale exceeds allowable limits.
Load on Scale is not stable.
Load Cell damage.
Lock Calibration Menu set to on.
LFT menu set to on.
Incorrect value for calibration mass.
Unit not set to on.
Menu has been locked.
Weight reading exceeds Power On Zero limit.
Weight reading below Power On Zero limit.
Weight reading exceeds Overload limit.
Weight reading below Underload limit.
Weight exceeds six digits. Display overflow.
Calibration data not present.
Batteries are discharged.
Calibration value outside allowable limits
Attempting to exit the menu with the LFT
setting ON and the security switch OFF.
Reference Weight too small. The weight
on the platform is too small to define a
valid reference weight.
Corrupted EEPROM data.
Check power cord connections. Make sure
power cord is plugged in into power outlet.
Check power source.
Replace batteries (T52P).
Service required.
Remove load on Scale.
Wait for load to become stable.
Service required.
Set Lock Calibration menu to off.
Set LFT menu to off.
Use correct calibration mass.
Enable unit in the Weighing Unit menu.
Refer to Weighing Unit on page 11.
Set selected menu to off in the Lock menu.
Lockout Switch on the circuit board may
need to be set to the off position.
Remove load from scale. Recalibrate scale.
Add load to scale. Recalibrate scale.
Reduce load on scale.
Add load to scale. Recalibrate scale.
Reduce load on scale.
Calibrate scale.
Replace batteries (T52P).
Use correct calibration weight.
Set the security switch to the ON position.
Use a greater weight for sample.
1-800-295-5510
uline.com
RECOMMENDATIONS
0621 IH-8105