Setup Menu - Uline Ohaus Ranger Count 3000 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Zero2:
Span1:
Span2:
Linearity1:
Linearity2:
Geographic
Adjustment:
End Calibration:
*Bold always represents factory default value.
ZERO2 [ZERO]
Initiates a second scale zero calibration.
SPAN1 [SPAN]
Initiates a span calibration procedure (zero and span).
A span calibration is important when initially setting up
the scale.
SPAN2 [SPAN2]
Initiates a second scale span calibration procedure.
LIN1 [LIN]
Initiates a linearity calibration procedure (zero,
mid-point and span).
LIN2 [LIN2]
Initiates a second scale linearity calibration procedure.
GEO [GEO]
Geographical Adjustment Factor (GEO) is used to adjust
the calibration based on the current location. Settings
from 0 to 31 are available with 12 being the default.
Determine the GEO factor that corresponds to your
location. (See table on page 15)
END CAL [END]
Advance to the next menu or return to the top of the
current menu.

SETUP MENU

Enter this menu to set scale parameters. Default settings
are in bold.
Reset
Dual Scale
Grad2
Power on unit
Zero Range
Auto Opt
Auto Tare
PAGE 9 OF 69
MENU SETTINGS CONTINUED
Perform
Perform
Perform
Perform
Perform
Set 0...12...31*
Exit menu
no, yes
off, on
1...9999
0.00005...0.5
auto, kg, g, lb, oz, lb:oz
off, on
on, off, on-acc
Beeper Opt
Beeper Signal
Beeper Key
Accumulation
End Setup
RESET [RESET]
Reset the Setup menu to factory defaults.
• NO = not reset
• YES = reset
DUAL SCALE [D.SCALE]
Set the status of the second scale input (Scale 2).
• OFF = disabled
• ON = enabled
CAP2 [CAP2]
Set the capacity of Scale 2.
• 1...9999
GRAD2 [GRAD2]
Set the readability of Scale 2.
• 0.00005...0.5
POWER ON UNIT [PWWR.UN]
Set the unit of measure displayed at startup.
• AUTO = last unit in use when turned off
• kg = kilograms
• g = grams
• lb = pounds
• oz = ounces
• lb:oz = pound ounces
ZERO RANGE [ZERO]
Set the percentage of scale capacity that may
be zeroed.
• 2% = zero up to 2 percent of capacity
• 10% = zero up to 10 percent of capacity
AUTO OPT [A.OPT]
Set the Automatic Optimization.
• OFF = disabled
• ON = APW automatically optimized
off, on
off, accept, under, over, under-over
off, on
off, auto, manual
Exit menu