Technical Tips
Which water should you use in your bath?
Use tap water with care. Water with a high lime content will cause scale build up and should
be avoided.
Distilled water and some types of de-ionised water may be used. Avoid ultra high purity de-
ionised waters.
Avoid using water with high levels of salts or iron. These will reduce the life of your bath
Regular water changing and frequent cleaning of your bath is needed to preserve the baths
corrosion resistance
Ensure you bath is stored dry.
Use care in placing other metallic items in the bath. Some metals (e.g. ferrous materials
such as iron filings and swarf) can cause an electro-chemical reaction leading to corrosion.
The product warranty may be affected by the use of inappropriate or corrosive liquids
7.
Troubleshooting
No lights or display
Check power source or mains switch
Display shows '' when shaking start button is pressed
Check that a shaking speed has been set by pressing the rpm button. Set the speed as
required.
The display keeps showing ''
The water temperature has exceeded the over temperature alarm set temperature. Check and
adjust the over temperature alarm temperature as required.
Display shows ''
The bath has been heating with no water or been allowed to boil dry. Check the water level,
and the control temperature. Switch the bath off and on to clear the alarm.
Display shows ''
The bath has tripped the fixed over temperature cut-out, an internal circuit breaker. It will need
to be checked by a qualified service technician. Contact your local VWR representative.
Display shows ''
Internal probe error. Contact your local VWR representative.
Display shows '...' flashing permanently.
System error. Contact your local VWR representative.
Display temperature is different from actual temperature
Use single point calibration () in section 3.5.4 to correct this.
There is no last digit on the display
Allow the bath to stabilize.
VWR shaking water bath Instruction Manual