2. DIRECTIONS FOR USE
Step 3:
PRIME
Press the
key again. Display will read MUTE WHEN DONE or BEEP WHEN
DONE. Press the + key to change to BEEP WHEN DONE or the
to MUTE WHEN DONE. If setting is as desired, do not press either key.
Step 4:
PRIME
Press the
key again. Display will read LITE ON or LITE OFF.
Press the + key to change to LITE ON or the
LITE OFF. If setting is as desired, do not press either key.
Step 5:
PRIME
Press the
key again. Display will read MAX RATE followed by NONE or a
maximum rate value such as 100 ml/hr. Press the
until the desired value is displayed. If setting is as desired, do not press either key.
Step 6:
PRIME
Press the
key again. Display will read MAX DOSE followed by NONE or a
maximum dose value such as 20 ml. Press the
until the desired value is displayed. If setting is as desired, do not press either key.
Step 7:
ON/OFF
Press the
NOTE: Once user has changed the desired user preference setting,
it is not necessary to view the remaining user preference settings.
Each change will be automatically saved as it is made.
NOTE: Pump will turn off automatically after 2 minutes in the User
Preference Settings if the user does not turn it off before then.
Infinity
Orange
Operator's Manual
®
®
key to turn pump off. Setting changes will be saved automatically.
-
key to change to
+
or
-
key to scroll between settings
+
-
or
key to scroll between settings
18
-
key to change