9. ADDITIONAL TECHNICAL INFORMATION

9. ADDITIONAL TECHNICAL INFORMATION
For any questions or additional information, refer
to the contact information on page 44
Infinity Orange Disposable Set Residual Fluid
When fully primed, the Infinity Orange Disposable
Set tubing contains approximately 5 mL of fluid.
As the Infinity Orange delivers fluid and empties
the bag, approximately 3.5 mL of air is drawn
into the tubing before the pump detects
that the bag is empty. Be sure to
account for this 1.5 ml of residual fluid
when filling the disposable bag sets,
especially for volume-sensitive users.
If you want to deliver 100 ml of
fluid, this is how you would calculate
how much fluid to place into the bag set.
Total Fluid
Desired
for Feeding
100 ml
Continue setting up the disposable bag set and pump according to CHAPTER
2: DIRECTIONS FOR USE, making sure to set the dose to 500 ml.
Electromagnetic Emissions/Interference
Electromagnetic emissions may affect the operation of any
electronic medical device, including enteral feeding pumps.
The Infinity Orange will not be affected by electromagnetic emissions in most
environments. However, some electromagnetic fields produced by personal
communication equipment, household appliances, or occupational tools may
cause electromagnetic interference (EMI) which can affect the pump.
Possible sources of electromagnetic interference with electronic medical devices include,
but are not limited to: cellular phones, cordless telephones, microwave ovens, anti-theft/
security systems, blenders, and high-powered tools (i.e. drills, saws, chain saws). If
electromagnetic emitting devices are operated within one yard/meter of the Infinity
Orange, the pump may automatically shut off and settings may return to their default
values. Check the pump regularly if operating near sources of electromagnetic emissions.
The Infinity Orange can safely be operated on commercial aircraft and
is designed in accordance with EN 60601-1-2, EN 60601-1-4 and RTCA
DO160D standards for electromagnetic emissions and immunity.
Infinity
Orange
Operator's Manual
®
®
Residual
+
=
Fluid
+
1.5 ml
=
3.5ml
Minimum Fluid
Required in
Bag Set
101.5 ml
36
1.5
Figure 9-1 Disposable
Set Residual Fluid
5ml
ml