Max Rate:
Rate range is 0.1 ml/hr up to 300 ml/hr. However,
the rate may be programmed to limit how
high the rate can be set. The limits are 50, 100,
150, 200, and 250 ml/hr. Rate is not limited if
user preference setting is programmed to NO
MAX. Pump default is NO MAX (Figure 2-25).
Max Dose:
Dose range is 0.1 ml up to 3000 ml. However, the
dose may be programmed to limit how high the
dose can be set. The limits are 10, 20, 30, 40, 50,
60, 70, 80, 90, 100, and 120 ml. Dose is not limited
if use preference setting is programmed to NO
MAX. Pump default is NO MAX (Figure 2-26).
To adjust these settings:
Key Sequence for "Adjusting User Preference Settings"
HIGH
LOW
MAX RATE
MAX RATE
Step 1:
With pump turned off, press and hold for 1.5 seconds the
OFF
keys at the same time. Pump will beep and then display the words
BEEP HIGH or BEEP LOW. Press the
change to LOW. Pump will give a triple beep when the
and setting will change. If setting is as desired, do not press either key.
Step 2:
PRIME
Press the
key. Display will read UNLK (unlocked) or LOCK (locked).
+
Press the
key to change to LOCK (locked) or the
UNLK (unlocked). If setting is as desired, do not press either key.
LOCK
UNLK
MAX DOSE
MAX DOSE
+
key to change to HIGH or the
17
2. DIRECTIONS FOR USE
Figure 2-25 MAX RATE Settings
Figure 2-26 MAX DOSE Settings
MUTE
BEEP
LITE OFF
+
and the
-
+
-
or
keys are pressed
-
key to change to
© Copyright 2020 Moog, Inc.
LITE ON
ON/
key to