Indications; To Stop Let Go; Max; Con Firm Dose And Food Type - Moog infinity orange Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

3. ALARMS, MESSAGES, AND INDICATIONS

Indications

TO STOP LET GO (Figure 3-10)
When pressing and holding the PRIME key during
manual prime, alarm will sound once, pump will
begin pumping at an approximate rate of 360 ml/ hr
and display will read "TO STOP LET GO"
To stop priming: Release the PRIME key.
MAX (Figure 3-11)
If a rate/dose limit has been programmed in the User
Preference Settings, pressing the
has been met will cause the pump to display MAX.
For example, if the User Preference Setting for
maximum rate has been programmed to 100 ml/
hr when the user attempts to program a rate higher
than 100 ml/hr the rate will stop increasing and
+
each time the
key is pressed the display will read
MAX. When released, display will read 100 ml/hr.
To correct the problem: Change limit in
User Preference Settings (Page 16).
CON FIRM DOSE AND FOOD TYPE (Figure 3-12)
All feeding parameters must be confirmed before
pump will begin running. This message will
RUN/PAUSE
display if the
viewing rate, dose, and food type parameters.
To correct the problem: View rate, dose,
and food type. Change settings where
required by using the
NOTE: The food type key allows users to select
whether they are delivering formula (when
selected, screen will read "FORM ULA") or human
milk (screen will read "HUMN MILK"). The pump
automatically calibrates the air-in-line sensor
sensitivity based on what food type setting is chosen.
Infinity
Orange
Operator's Manual
®
®
+
key when that limit
key is pressed prior to
+
-
or
key.
24

Figure 3-10 Priming Indication

Figure 3-11 MAX Indication

Figure 3-12 CON FIRM DOSE
AND FOOD TYPE Indication