5. CLEANING
WARNING: To avoid electrical shock, never clean pump or Infinity Orange AC
Adapter/Charger with charger plugged into an outlet or pump turned on.
WARNING: Make sure the Infinity Orange AC Adapter/Charger is
completely dry before plugging into an electrical outlet.
To clean the Infinity Orange enteral feeding pump:
Pump may be cleaned with warm soapy water
(standard dish soap) and a nonabrasive sponge or
soft cloth. Use a cotton swab to clean pathways
of cassette receptacle and to remove orange
silicon residue from pump wheel roller pins.
OR - Apply one of the following solutions for
approximately 10 minutes, then wipe pump
clean with a damp cloth or sponge:
4 5% bleach and water solution
4 Multipurpose household disinfectant cleaner
Rinse pump by holding under
a stream of warm water. Then, dry with a clean cloth.
NOTE: Clean pump as needed. If the pump is used on more
than one patient, clean between each patient.
NOTE: Avoid harsh cleaners/disinfectants. They may cause damage to
pump surfaces and may affect pump functionality. Moog has tested the
following disinfectants for acceptability for cleaning/disinfecting the Infinity
Orange pump. These cleaning products are categorized below:
Acceptable
Metrex CaviCide
5% Bleach and Water Solution
Warm Soapy Water (Standard Dish Soap)
To clean the Carrying Packs:
The Infinity brand packs are machine-washable. Use cold water
and gentle cycle, whenever possible. Hang to dry.
To clean the AC Adapter/Charger:
Infinity Orange AC Adapter/Charger normally does not require cleaning. When
desired, a dry or slightly damp cloth may be used to clean the outside surface
of AC Adapter/Charger while it is disconnected from the wall outlet.
Unacceptable
Ball Ruthless
Wex-Cide 128
Alcohol
Vesphene Ilse
29
5. CLEANING
Figure 5-1 Rinsing Pump
Under Stream of Water
© Copyright 2020 Moog, Inc.