Contraindications
Stryker promotes the assessment of each patient and appropriate usage by the operator. Contraindications for this
product are for patients with the conditions specified below.
Heating contraindications include:
•
Application to a body surface with compromised blood flow (Ischemia, area under pressure, arterial insufficiency)
•
Application to a patient with an increased tendency to bleed (aggravates potential for hemorrhage)
•
Application to a body surface with possible malignancy (tissue metabolism is increased and therefore, the growth
potential of the malignant tissues)
•
Treatment of hematoma within the first 24 to 48 hours (potential for re-bleeding and hemorrhage). Recent sprain or
fracture (acute inflammatory response).
•
In combination with topical solutions whose toxicity may be affected by the application of heat
•
In combination with other heat sources
Cooling contraindications include:
•
Application to a body surface with compromised blood flow (Ischemia, area under pressure, arterial insufficiency)
•
Application to body surface with known vascular impairment such as frostbite, arteriosclerosis or ischemia
•
Application to body surface in patients with hypersensitivity to cold, such as people with Raynaud's phenomenon,
cold urticaria, cryoglobulinemia, and paraxysmal cold hemoglobinuria
•
Application to body surface in patients with impaired sensation
•
In combination with topical solutions whose toxicity may be affected by the application of cold
Specifications
Models
Temperature setpoints Fahrenheit
Temperature setpoints Celsius
Therapy duration
Classification
Size
Weight
Reservoir empty
Reservoir filled with water to the
heating level
Maximum reservoir capacity
Flow rate, minimum with pad attached
Average temperature accuracy
Maximum contact surface temperature
High limit safety temperature
Power cord
Dual hose
Current leakage, maximum
Ground resistance, maximum
Electrical requirements
www.stryker.com
Introduction
TP700 / TP700C
107° F, 100° F, 95° F, 50° F
42° C, 38° C, 35° C, 10° C
20 or 30 minutes or continuous cycle
Class I equipment with Type BF applied part suitable for continuous
operation. Not classified for protection against ingress of liquid. Not
classified for use in the presence of flammable anesthetics.
11.5 in. x 8 in. x 8 in.
6.5 lb
9 lb
93 oz
9 gph
±2° F at 107° F
107° F
110° F to 120° F
International (harmonized) 3 wire cord set
10 ft
100 µA
0.5 mΩ
Voltage
Frequency:
Current in amperes
101515 REV E
29.2 cm x 20.3 cm x 20.3 cm
2.9 kg
4.0 kg
2.75 l
34 lph
±1° C to 42° C
42° C
43.3° C to 49° C
305 cm
120 VAC ±10%
60 Hz
3.1
English
EN
1-5