Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Note:
U
is the a.c. mains voltage prior to applications of the test level.
T
Recommended separation distances between portable and mobile RF communications equipment and the T/
The T/Pump system is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the T/Pump system can help prevent electromagnetic interferences by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
T/Pump system as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
The T/Pump system is suited for use in the electromagnetic environment specified below. The customer or the user
of the T/Pump system should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
www.stryker.com
EMC Information
3 A/m
Pump.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d=1,2√P
0,12
0,38
1,2
3,8
12
Compliance Level
101515 REV E
3 A/m
m
80 MHz to 800 MHz
d=1,2√P
0,12
0,38
1,2
3,8
12
Electromagnetic Environment - Guidance
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
800 MHz to 2,5 GHz
d=2,3√P
0,23
0,73
2,3
7,3
23
1-19
English
EN