SAFETY
The instruments meet the requirements for double insula-
tion to IEC 1010-1 (1995), EN 61010-1 (1995) to Category
III*, 300 Volts phase to earth (ground) and 440 Volts phase
to phase, without the need for separately fused test leads.
If required, fused test leads are available as an optional
accessory.
* Relates to the transient over-voltages likely to be met in
fixed wiring installations.
Complies with the following parts of EN61557, Electrical
safety in low voltage systems up to 1000V a.c. and 1500V
d.c. – Equipment for testing, measuring or monitoring of
protective measures:-
Part 1 - General requirements
Part 2 - Insulation resistance
Part 4 - Resistance of earth connection and equi-
potential bonding
FUSE
500mA 500V, 32x 6mm Ceramic HBC 10kA minimum.
E.M.C.
The instruments meet EN61326-1

POWER SUPPLY

Battery Type:
6x1,5V Alkaline cells
IEC LR6 type or 1.2V NiCd
re-chargeable cells.
Battery Life (typical): 2100 5-sec 1kV insulation tests
3200 5-sec 500V insulation tests
4000 5-sec 250V insulation tests
2700 5-sec continuity tests
4700 5-sec kΩ tests
ENVIRONMENTAL CONDITIONS
Operating range:
Operating humidity:
Storage temperature range: -25 to +65°C
Calibration Temperature:
Maximum altitude:
Dust and water protection:
Temperature coefficient:
WEIGHT:
DIMENSIONS:
CLEANING:
-5 to +40°C
90% RH at 40°C max.
+20°C
2000m
IP54
<0,1% per °C
742g
110mm x 220mm x 45mm
Wipe with a clean cloth
dampened with soapy
water or Isopropyl Alcohol
(IPA)