(0,01kΩ to 10000kΩ) make the resistance range
very useful for general purpose testing.
Voltage Tests (V)
If >1V a.c. or d.c. is present at the terminals the
measured voltage is indicated on the display. The
voltage display will function within specification
even if the fuse has blown.
If the voltmeter operation is in question, test the
voltmeter on a known source.
1. Set the selector switch to V.
2. Connect the test leads.
3. After a short settling time, the reading will be
displayed automatically.
To view the frequency of the a.c. V being measure d
p ress the
k e y. The frequency will be displayed
in the range 16Hz-460Hz. To view a.c. V press the
key again.
Millivolt Tests(mV)
The measured a.c. or d.c. voltage is indicated on
the display.
1. Set the selector switch to mV.
2. Select either a.c. or d.c. mV using the
key.
3. Connect the test leads.
4. After a short settling time, the reading will be
displayed automatically.
Note: Live circuit warning does not function on mV
range.
Zeroing of d.c. mV (no a.c. mV zero facility)
To zero the d.c. mV range, short the leads togeth-
er in the d.c. mV position, wait for the reading to
settle and then press the TEST button. Up to
9,9mV can be zeroed on the d.c. mV range. The
symbol
will appear to indicate the zero has
been adjusted.
1. Select the d.c. mV range.
2. Short the test leads together.
3. When the reading has stabilised, press the
TEST button. The zero offset symbol
appear.
4. To release the zero offset press the TEST but-
ton again.
To view the frequency of the a.c. mV being mea-
sured press the
displayed in the range 16Hz-460Hz. To view a.c.
key. The frequency will be
will
11