Maintenance and Care:

The Infinity XR2 Base Unit should be returned to its storage case after each use.
NOTE: Ensure all components have completely dried after cleaning before returning to case.
If unit is dropped or mishandled, it should be inspected for damage.
(REF Inspection section of this instruction manual) If damage occurs, do not
use; return the complete device immediately to Integra LifeSciences
for inspection.
Inspection of Components
A routine inspection of the components of the MAYFIELD Infinity XR2 Base Unit should be made
before each and every procedure to assist in keeping it in good functional condition and to avoid
problems the day of surgery. This check should include the following:
1. Check to see that all the components of the base unit are available for assembly. Use the
Inspection section of this Instruction Manual for a complete list of components. ALL
components must be available and ready for assembly or the base unit should not be
used.
2. Check the adjustment of the base locking handles by following these steps.
a. Attach the base unit to an OR table and lock to the side rails.
b. Raise the Crossbar to be even with the top of the OR table, locking the handles.
c. With your hands on either ends of the Crossbar, push down on the Crossbar attempting
to move it. If with minimal force, the Crossbar moves, adjustment to increase the
locking power of the Locking Handles should be made as outlined in this manual.
d. If the handles are too hard to close, adjustment to decrease the locking power of the
Locking Handles should be made as outlined in this manual.
3. With the Locking Handles unlocked and all the components attached but loose, raise and
lower the base unit, rotating all joints. All should freely rotate without any "ratcheting"
noise (starburst teeth hitting together as rotated). If this ratcheting noise is heard, loosen
the joint that is causing the noise.
4. Check the other components for function
a. Again, with the base unit Crossbar raised to be level with the top of the OR table, lock
the locking handles.
b. Lock the link arm locking knobs
c. Verify that all starbursts are fully engaged and locked
d. With these knobs locked, the total base unit should be locked into place and no
movement should be seen.
CAUTION!
Page 11