COMPLIANCE
This component has been tested according to ISO 10328 standard to two
million load cycles. Depending on the amputee's activity this
corresponds to a duration of use of two to three years. We recommend
carrying out regular yearly safety checks
ISO
- "P" - "m"kg *)
*)
Body mass limit not to be exceeded!
For speci c conditions and limitations of use see
manufacturer's written instructions on intended use!
have been adapted adequately on the basis of the specified load level.
Category
1
2
3
4
5
6
7
8
9
• This manual is intended for use by a certified prosthetist.
9
In the standard mentioned, test levels
(P) are assigned to a certain maximal
body masses (m in kg). In some cases,
which are marked with, no test level is
assigned to the product related maximal
body mass. In these cases, the test loads
Category Össur high activity
Weight (Kg)
52
59
68
77
88
100
116
130
147
Lable text
P3
ISO 10328 -
P3
ISO 10328 -
P4
ISO 10328 -
P4
ISO 10328 -
P4
ISO 10328 -
P5
ISO 10328 -
P5
ISO 10328 -
P6
ISO 10328 -
P6
ISO 10328 -
P3
52
kg
P3
59
kg
P4
68
kg
P4
77
kg
P5
88
kg
P5
100
kg
P6
116
kg
P6
130
kg
P6
147
kg