• If you store the device and cuff, do not place heavy ob-
jects on the device and cuff. Remove the batteries. The
cuff line should not be bent sharply.

10. Technical details

Model no.
BC 44
Measurement
Oscillometric, non-invasive blood pres-
method
sure measurement on the wrist
Measurement
Cuff pressure 0 – 300 mmHg,
range
systolic 60 – 260 mmHg,
diastolic 40 –199 mmHg,
Pulse 40 –180 beats/minute
Display accuracy Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Measurement
Max. permissible standard deviation
inaccuracy
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
Dimensions
L 87,8 mm x W 65,8 mm x H 28,6 mm
Weight
Approx. 98 g
(without batteries, with cuff)
Cuff size
140 to 195 mm
Permissible oper-
+ 10 °C to + 40 °C, ≤ 90 % relative air
ating conditions
humidity (non-condensing)
Permissible stor-
- 20 °C to + 55 °C, ≤ 90 % relative air hu-
age conditions
midity, 800 –1050 hPa ambient pressure
Power supply
2 x 1,5 V
Battery life
For approx. 180 measurements, depend-
ing on the blood pressure level and/or
pump pressure
Classification
Internal supply, continuous operation,
type BF applied part, IPX0, no AP or
APG
The serial number is located on the device or in the battery
compartment.
Technical information is subject to change without notifica-
tion to allow for updates.
• This unit is in line with European Standard EN 60601-1-2
(In accordance with CISPR 11, IEC 61000-4-2,
IEC 61000-4-3 and IEC 61000-4-8) and is subject to par-
ticular precautions with regard to electromagnetic com-
patibility (EMC). Please note that portable and mobile HF
communication systems may interfere with this unit.
• This device is in line with the EU Medical Devices Direc-
tive 93/42/EEC, the "Medizinproduktegesetz" (German
Medical Devices Act) and the standards EN 1060-1
(non-invasive sphygmomanometers, Part 1: General
requirements), EN 1060-3 (non-invasive sphygmoma-
nometers, Part 3: Supplementary requirements for
25
AAA batteries