Error Messages/Troubleshooting; Technical Details; Cleaning And Storing The Instrument - Beurer BC 19 Mode D'emploi

Masquer les pouces Voir aussi pour BC 19:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 22

7. error messages/troubleshooting

In case of faults, the ‚Er_' message appears in the
display.
Error messages can occur when
• inflation takes longer than 3 minutes (Er6 appears in
the display),
• the cuff is too tight or too loose (Er3 or Er4 appears in
the display),
• the blood-pressure values are unusually high or low
(Er1 or Er2 appears in the display),
• the inflation pressure is higher than 300 mmHg (Er5
appears in the display),
• there is a system or device error (Er0, Er7, Er8, Er9,
ErA or ErB appears in the display)
• the batteries are almost flat; the batteries must be
replaced.
In the above cases, you must repeat the measurement.
Make sure that the cuff tube is properly inserted and that
you do not move or talk.
Re-insert the batteries if necessary, or else replace
them.

8. cleaning and storing the instrument

• Clean your blood pressure computer carefully using
only a slightly moistened cloth.
• Do not use any detergents or solvents.
All manuals and user guides at all-guides.com
• Never hold the instrument under water as otherwise
• Never place any heavy objects on the instrument.

9. Technical details

Model no.
Measurement
method:
Measurement range
Precision of
pressure reading
Measurement
uncertainty
Memory
Dimensions
Weight
Cuff size
Admissible operating
temperature
19
liquid can penetrate and damage the instrument.
BC 19
Oscillating, non-invasive blood
pressure measurement on the
wrist
Pressure 0-300 mmHg, heart
rate 40-180 beats/minute
Systolic ± 3 mmHg/
diastolic ± 3 mmHg
pulse ± 5% of displayed value
Max. admissible standard
deviation according to clinical
examination: systolic 8 mmHg/
diastolic 8 mmHg
2 x 60 storable data lines
L 85 x W 70 x H 31 mm
Approx. 140 g
For a wrist circumference of ap-
prox. 140 to 195 mm
+5 °C to +40 °C, 40-80 %
relative humidity