Error Messages/Troubleshooting; Cleaning And Storing The Device And Cuff - Beurer BC 44 Mode D'emploi

Masquer les pouces Voir aussi pour BC 44:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 28
Blood pressure
Systole
value category
(in mmHg)
High normal
130 – 139 85 – 89
Normal
120 – 129 80 – 84
Optimal
< 120
Source: WHO, 1999 (World Health Organization)

8. Error messages/troubleshooting

In case of faults, the
_ message appears in the display.
• Error messages may appear if:
• systolic or diastolic pressure could not be measured (
or
appears on the display)
• systolic or diastolic pressure was outside the measure-
ment range ("Hi" or "Lo" appears on the display)
• the cuff is fastened too tightly or loosely (
pears on the display)
• the pump pressure is higher than 300 mmHg (
pears on the display)
• pumping up takes longer than 160 seconds (
on the display)
• there is a system or device error (
appears on the display)
• the batteries are almost empty
Diastole
Action
(in mmHg)
regular monito-
ring by doctor
self-monitoring
< 80
self-monitoring
or
ap-
ap-
appears
,
,
or
.
In such cases, repeat the measurement. Ensure that you do
not move or speak during the measurement. If necessary,
reinsert or replace the batteries.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic)
lie outside the limits specified in the section "Technical
specifications", the technical alarm will appear on the dis-
play indicating either "Hi" or "Lo". In such cases, you should
seek medical assistance and check the accuracy of your
procedure.
The limit values for the technical alarm are factory set and
cannot be adjusted or deactivated. These alarm limit val-
ues are accorded second priority under the standard IEC
60601-1-8.
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear auto-
matically after about 8 seconds.

9. Cleaning and storing the device and cuff

• Clean the device and cuff carefully using a slightly damp
cloth only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under
water, as this can cause liquid to enter and damage the
device and cuff.
24

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

65905B002kfzw1k141919

Table des Matières