• Before the initial blood pressure measurement, make sure
always to rest for about 5 minutes.
• Furthermore, if you want to take several measurements in
succession, make sure always to wait for at least 1 minute
between the individual measurements.
• Repeat the measurement if you are unsure of the meas-
ured value.
• The measurements taken by you are for your information
only – they are not a substitute for a medical examination!
Discuss the measurements with your doctor, and never
base any medical decisions on them (e.g. medicines and
their administration)!
• Using the blood pressure monitor outside your home en-
vironment or whilst on the move (e.g. whilst travelling in a
car, ambulance or helicopter, or whilst undertaking physi-
cal activity such as playing sport) can influence the meas-
urement accuracy and cause incorrect measurements.
• Do not use the blood pressure monitor on newborns or pa-
tients with preeclampsia. We recommend consulting a doctor
before using the blood pressure monitor during pregnancy.
• This device is not intended for use by people (including
children) with restricted physical, sensory or mental skills
or a lack of experience and/or a lack of knowledge, un-
less they are supervised by a person who is responsible
for their safety or are instructed by such a person in how
to use the device. Supervise children around the device to
ensure they do not play with it.
• In the case of restricted circulation on the arm as a result
of chronic or acute vascular diseases (including vascular
constriction), the accuracy of the wrist measurement is
limited. In this case you should avoid using an upper arm
blood pressure monitor.
• Cardiovascular diseases may lead to incorrect measure-
ments or have a detrimental effect on measurement ac-
curacy. The same also applies to very low blood pressure,
diabetes, circulatory disorders and arrhythmias as well as
chills or shaking.
• The blood pressure monitor must not be used in connec-
tion with a high-frequency surgical unit.
• Only use the unit on people who have the specified wrist
measurement for the device.
• Please note that when inflating, the functions of the limb
in question may be impaired.
• During the blood pressure measurement, blood circulation
must not be stopped for an unnecessarily long time. If the
device malfunctions, remove the cuff from the arm.
• Do not allow sustained pressure in the cuff or frequent
measurements. The resulting restriction of the blood flow
may cause injury.
• Ensure that the cuff is not placed on an arm in which the ar-
teries or veins are undergoing medical treatment, e.g. intra-
vascular access or therapy, or an arteriovenous (AV) shunt.
• Do not use the cuff on people who have undergone a
mastectomy.
17