14. ELECTROMAGNETIC COMPATIBILITY (EMC) TABLES
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The CEFAR EASY Electrotherapy System is intended for use in the electromagnetic environment deified below. The customer
or the user of the CEFAR EASY Electrotherapy System should assure that it is used in such an environment
Emissions Tests
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Not Applicable - Battery
IEC 61000-3-2
powered
Voltage fluctuations
Not Applicable - Battery
IEC 61000-3-3
powered
Compliance
The CEFAR EASY Electrotherapy System uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic equipment.
CEFAR EASY Electrotherapy System is suitable for use in all
establishments, including domestic establishments and those
directly connected to the public low voltage power supply network
that supplies buildings used for domestic purposes
Electromagnetic Environment - Guidance