THE STIMULATION FEELS WEAK OR NOT AT ALL
• Check if the batteries need to be replaced.
· Electrodes are too old and need replacement.
THE BROKEN CIRCUIT SYMBOL IS SHOWN ON THE DISPLAY SYMBOL
The broken circuit symbol indicates that the resistance is too high, or that a cable is broken.
· A too high resistance can be caused by a bad connection between the electrodes and your
skin, or that the electrodes need to be replaced.
· A cable breakage can be checked by pressing the cable´s pins against one another while in-
creasing the amplitude for the corresponding channel to 11 mA. If the amplitude now drops
to 0.0 mA and
starts flashing, the cable needs to be replaced.
Note! Never increase the amplitude above 20 mA when you check for cable breaks, since this
can damage the stimulator.
THE STIMULATOR IS NOT WORKING
If the display to the left appears when you start the stimulator, it means that
the stimulator is broken and needs to be replaced. NOTE! Do not use the stim-
ulator - contact your Cefar dealer.
10. TECHNICAL DATA
CEFAR EASY is an electrical stimulator intended for Pain Relief and Massage.
It delivers a constant current of 60 mA through a load of 1000 Ω.
INFORMATION RELATED TO ELECTROMAGNETIC COMPATIBILITY (EMC).
CEFAR EASY is designed to be used in typical domestic or commercial environments and ap-
proved according to the EMC safety standard of EN 60601-1-2.
CEFAR EASY emits very low levels in the radio frequency (RF) interval. Therefore it is not likely
to cause any interference in nearby electronic equipment (radios, computers, telephones etc.).
CEFAR EASY is designed to withstand foreseeable disturbances originating from electrostat-
ic discharges, mains supply magnetic fields and radio frequency transmitters (such as mobile
telephones).