- Sterilization -
<Validated Sterilization Method>
Standard followed
Sterilization method
Sterilization conditions
* According to the Guideline for Disinfection and Sterilization in
Healthcare Facilities, 2008 (Update: May 2019)
Do not sterilize using low-temperature hydrogen
peroxide gas plasma. It may cause discoloration and/or
deformation.
With a high-pressure steam sterilizer that has been
confirmed to function properly, the validation of
sterilization is verified by a medical institution,
sterilization conditions with a proven effectiveness
shall be applied.
Pre-vacuum steam sterilization
(Autoclave sterilization)
Sterilization
temp.
132 °C / 270 °F*
134 °C / 273 °F*
-19-
ISO 17665-1
Retention
Drying time
time
4 minutes
20 minutes or
or more
3 minutes
20 minutes or
or more
more
more