Sterilization equipment should be cleaned and
maintained in accordance with standardized operating
procedures specified by its manufacturer.
Clean and sterilize the clip applier according to the
clip applier instruction manual.
This clip and instruments must be cleaned to reduce
bioburden before first sterilization and resterilization.
Do not use chlorine-based solutions when cleaning this
clip. This could cause metallic corrosion.
1) Cleaning
Put the clip in the designated Sugita Clip Case or
Sugita Clip Tray (hereinafter this is called "clip case or
tray") and close the lid.
<In case of using Ultrasonic cleaner>
(1) Fully immerse the case (tray) in the water (tap
water) of the washing tank.
(2) Ultrasonic cleaning
Cleaning time: 15 minutes or more
Detergent: Neutral enzyme type detergent (Diluted
-15-